We started a cycle and the monitor immediately displayed cycle complete? Why do we have to add chemical? The cycle will not start and the monitor shows a minimum reservoir temperature not achieved? I have noticed that we have to add chemical more often than normal what could be the problem? How do I send an order to Custom Ultrasonics for parts?
We started a cycle and the monitor immediately displayed cycle complete?
Notate the type of error that is displayed and contact Custom Ultrasonics. This will help Custom Ultrasonics to better diagnose the problem.
Back to Top
Why do we have to add chemical?
The Custom Ultrasonics System 83 disposes 3 to 4 ounces of chemical down the drain each cycle to prevent dilution of the chemical being transferred back to the chemical reservoir.
Back to Top
The cycle will not start and the monitor shows a minimum reservoir temperature not achieved?
If the chemical you are using needs to achieve a certain temperature, per the manufacturer’s label, the System 83 will not begin the cycle until the chemical reservoir has achieved the minimum operating temperature.
Back to Top
I have noticed that we have to add chemical more often than normal what could be the problem?
There may be debris in the drain and/or residual drain solenoids valve and they need to be cleaned immediately. Be sure to complete your Preventative Maintenance on a quarterly basis to ensure that your System 83 performs up to specification.
Back to Top